Specifications

Menu Name Function Description GPIB Mnemonic
AUTOCAL
ASSURANCE
LIMITS
Displays in MS4623X with a 4-port AutoCal box.
DIRECTIVITY
DC - 3 GHZ
40.0 DB
3-6GHZ
35.0 DB
Changes the value of AutoCal directivity limits.
The values are defaulted to
40.0 dB for DC to 3 GHz and
35.0 dB for 3 to 6 GHz.
ACADIR1;
ACADIR1?;
ACADIR2;
ACADIR2?
SOURCE MATCH
DC - 3 GHZ
38.0 DB
3-6GHZ
35.0 DB
Changes the value of AutoCal source match limits.
The values are defaulted to
38.0 dB for DC to 3 GHz and
35.0 dB for 3 to 6 GHz.
ACASRC1;
ACASRC1?;
ACASRC2;
ACASRC2?
REFLECTION TRACKING
DC - 3 GHZ
0.150 DB
3-6GHZ
0.200 DB
Changes the value of AutoCal reflection tracking
limits. The values are defaulted to
0.150 dB for DC to 3 GHz and
0.200 dB for 3 to 6 GHz.
ACARET1;
ACARET1?;
ACARET2;
ACARET2?
MORE Changes the soft-key display to the appropriate
AUTOCAL ASSURANCE LIMITS 2 menu.
None
RETURN Changes the soft-key display to the AUTOCAL
ASSURANCE menu.
None
AUTOCAL
ASSURANCE
LIMITS 2
Displays in MS4623X with a 4-port AutoCal box.
LOAD MATCH
DC - 3 GHZ
40.0 DB
3-6GHZ
35.0 DB
Changes the values of AutoCal load match limits.
The values are defaulted to
40.0 dB for DC to 3 GHz and
35.0 dB for 3 to 6 GHz.
ACALM1; ACALM1?;
ACALM2; ACALM2?
TRANSMISSION
TRACKING
DC - 3 GHZ
0.150 DB
3-6GHZ
0.200 DB
Changes the values of AutoCal transmission
tracking limits. The values are defaulted to
0.150 dB for DC to 3 GHz and
0.200 dB for 3 to 6 GHz.
ACATRT1;
ACATRT1?;
ACATRT2;
ACATRT2?
MORE Changes the soft-key display to the appropriate
AUTOCAL ASSURANCE LIMITS menu.
None
RETURN Changes the soft-key display to the AUTOCAL
ASSURANCE menu.
None
UTILITY FRONT PANEL
KEY MENUS OPERATION
4-90 MS462XX OM