Specifications

Menu Name Function Description GPIB Mnemonic
4 PORT
CALIBRATION
(SOLT/SOLR)
CONNECT A
RECIPROCAL
BETWEEN PORT
X AND Y
When a 3 Port calibration is already complete,
Ports X and Y can be:
Port 1 and Port 4
Port 2 and Port 4 or
Port 3 and Port 4
When a 3 Port calibration is not complete,
Ports X and Y can be:
Port 1 and Port 3
Port 1 and Port 4
Port 2 and Port 3
Port 2 and Port 4 or
Port 3 and Port 4
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
(SOLT/SOLR)
LEAVE A RECIPROCAL
CONNECTED
BETWEEN PORT
X AND Y
When a 3 Port calibration is already complete,
Ports X and Y can be:
Port 1 and Port 4
Port 2 and Port 4 or
Port 3 and Port 4
When a 3 Port calibration is not complete,
Ports X and Y can be:
Port 1 and Port 3
Port 1 and Port 4
Port 2 and Port 3
Port 2 and Port 4 or
Port 3 and Port 4
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
(SOLT/SOLR)
CONNECT A THROUGH
LINE BETWEEN PORT
X AND Y
When a 3 Port calibration is already complete,
Ports X and Y can be:
Port 1 and Port 4
Port 2 and Port 4 or
Port 3 and Port 4
When a 3 Port calibration is not complete,
Ports X and Y can be:
Port 1 and Port 3
Port 1 and Port 4
Port 2 and Port 3
Port 2 and Port 4 or
Port 3 and Port 4
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
CONNECT
CALIBRATION DEVICE
None
PORT 4:
OPEN/SHORT/
BROADBAND
Connect specified open, short, or broadband load to
Port 4, then press the MEASURE DEVICES soft-key.
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
FRONT PANEL CAL
OPERATION KEY MENUS
MS462XX OM 4-131