Specifications

Menu Name Function Description GPIB Mnemonic
4 PORT
CALIBRATION
CONNECT
CALIBRATION DEVICE
None
PORT 3: OPEN
PORT 4: SHORT
Connect specified open to Port 3 and short to Port 4,
then press the MEASURE DEVICES soft-key.
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
CONNECT
CALIBRATION DEVICE
None
PORT 3: SHORT
PORT 4: OPEN
Connect specified short to Port 3 and open to Port 4,
then press the MEASURE DEVICES soft-key.
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
CONNECT
CALIBRATION DEVICE
None
PORT 3: BROADBAND
PORT 4: BROADBAND
Connect specified broadband load to Ports 3 and 4,
then press the MEASURE DEVICES soft-key.
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
CONNECT
CALIBRATION DEVICE
None
PORT 4:
OPEN/SHORT/
BROADBAND
Connect specified open, short, or broadband load to
Port 4, then press the MEASURE DEVICES soft-key.
None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
4 PORT
CALIBRATION
MEASURING... None
RESTART
MEASUREMENT
Changes the soft-key display to the beginning 4
PORT CALIBRATION menu.
None
4 PORT
CALIBRATION
REMEASURE Changes the soft-key display to the previous 4 PORT
CALIBRATION menu.
None
NEXT CAL STEP Changes the soft-key display to next 4 PORT
CALIBRATION menu.
NCS
4 PORT
CALIBRATION
CONNECT ISOLATION
DEVICES TO ALL
PORTS
Connect appropriate isolation devices to all test ports. None
MEASURE DEVICE(S) Changes the soft-key display to the next 4 PORT
CALIBRATION menu.
TCD
CAL FRONT PANEL
KEY MENUS OPERATION
4-132 MS462XX OM