Specifications

Menu Name Function Description GPIB Mnemonic
CALIBRATION
SEQUENCE2
CONNECT THE
RECIPROCAL
XXXXXX mm
BETWEEN
TEST PORTS
None
MEASURE FORWARD
((NO) CAL EXISTS)
Measures calibration data in the forward direction,
then changes the soft-key display to the
CALIBRATION SEQUENCE CONTINUED menu.
TC1
MEASURE REVERSE
((NO) CAL EXISTS)
Measures calibration data in the reverse direction,
then changes the soft-key display to the
CALIBRATION SEQUENCE CONTINUED menu.
TC2
MEASURE DEVICE(S) Measures the throughline calibration data in both the
forward and reverse directions, then changes the
soft-key display to the CALIBRATION SEQUENCE
CONTINUED menu.
TCD
CALIBRATION
SEQUENCE3
CONNECT THE
THROUGHLINE
XXXXXX mm
BETWEEN
TEST PORTS
None
MEASURE FORWARD
((NO) CAL EXISTS)
Measures calibration data in the forward direction,
then changes the soft-key display to the
CALIBRATION SEQUENCE CONTINUED menu.
TC1
MEASURE REVERSE
((NO) CAL EXISTS)
Measures calibration data in the reverse direction,
then changes the soft-key display to the
CALIBRATION SEQUENCE CONTINUED menu.
TC2
MEASURE DEVICE(S) Measure the throughline calibration data in both the
forward and reverse directions, then changes the
soft-key display to the 2 PORT CALIBRATION menu.
TCD
CALIBRATION
SEQUENCE
CONTINUED
REMEASURE Changes the soft-key display to the previous
CALIBRATION SEQUENCE menu.
None
NEXT CAL STEP Changes the soft-key display to the next
CALIBRATION SEQUENCE menu.
None
CALIBRATION
SEQUENCE
COMPLETED
PRESS
<SAVE/RECALL> TO
STORE CAL DATA ON
DISK
OR
PRESS <ENTER> TO
PROCEED
Press the Save/Recall key to store the calibration
data on the hard or floppy disk.
Alternatively, press the Enter key to proceed.
SAVE “C:\*.CAL”
SAVE “A:\*.CAL”
CHANGE 3 PORT
CAL METHOD
24-TERM
SOLT (NOT FOR
WAVEGUIDE)
or with Option 24:
SOLT/SOLR (NOT FOR
WAVEGUIDE)
This option, and the ones below, provide for selecting
the method (procedure) to be used for calibration.
SOLT
TRX Selects TRX method. TRX
2-PATH 3-PORT (S11,
S21, S31)
Selects 2-Path, 3-Port method.
NEXT CAL STEP Changes the soft-key display to the BEGIN 3 PORT
CALIBRATION menu.
None
CAL FRONT PANEL
KEY MENUS OPERATION