Specifications

Menu Name Function Description GPIB Mnemonic
FULL TERM CAL
(FULL -PORT CAL)
PORT 1 ON
PORT 2 ON
Displayed when a 12-term cal exists.
Toggles port 1 selection on or off.
Toggles port 2 selection on or off.
FXP1T0; FXP1T1;
FXP1T?; FXP2T0;
FXP2T1; FXP2T?
PORT 1 ON
PORT 2 ON
PORT 3 ON
Displayed when a 24-term cal exists.
Toggles port 1 selection on or off.
Toggles port 2 selection on or off.
Toggles port 3 selection on or off.
FXP1T0; FXP1T1;
FXP1T?; FXP2T0;
FXP2T1; FXP2T?;
FXP3T0; FXP3T1;
FXP3T?
PORT 1 ON
PORT 2 ON
PORT 3 ON
PORT 4 ON
Displayed when a 40-term cal exists.
Toggles port 1 selection on or off.
Toggles port 2 selection on or off.
Toggles port 3 selection on or off.
Toggles port 4 selection on or off.
FXP1T0; FXP1T1;
FXP1T?; FXP2T0;
FXP2T1; FXP2T?;
FXP3T0; FXP3T1;
FXP3T?; FXP4T0;
FXP4T1; FXP4T?
RETURN Changes the soft-key display to the FLEXIBLE CAL
menu.
None
FREQUENCY
TRANSLATION
GROUP DELAY
CALIBRATION
APPLY CAL ON/OFF
(CAL EXIST/NO CAL
EXIST)
Turn on or off Frequency Translation Group Delay
correction.
FTGDC0; FTGDC1,
FTGDCX?;
FTGDCDONE?
PERFORM CAL Changes the soft-key display to the next soft-key
menu in calibration sequence. Refer to Appendix D
for menu flow.
BEGFTGD
HELP ON/OFF Displays an informational help screen. HELP1; HELP0;
HELPX?
FREQUENCY
TRANSLATION
GROUP DELAY
CALIBRATION 2
CONNECT A
THROUGHLINE
BETWEEN PORTS 1
AND 2
None
ABORT CAL Aborts the calibration. ABORTCAL
MEASURE DEVICE(S) Measures the device(s). None
FREQUENCY
TRANSLATION
GROUP DELAY
CALIBRATION 3
RESTART
MEASUREMENT
Restarts the measurement.
ABORT CAL Aborts the calibration. ABORTCAL
MEASURING
THROUGHLINE
BETWEEN PORTS 1
AND 2 ...
None
FREQUENCY
TRANSLATION
GROUP DELAY
CALIBRATION 4
SAVE CAL DATA Changes the soft-key display to the SAVE menu. SAVE *.cal
RESTART
MEASUREMENT
Restarts the measurement. None
EXIT Exit calibration menu. None
FRONT PANEL CAL
OPERATION KEY MENUS
MS462XX OM 4-159