Specifications

Menu Name Function Description GPIB Mnemonic
PORT X
OPEN DEVICE
ENTER THE
CAPACITANCE
COEFFICIENTS
None
TERM 1-C0
P/MXXX.XXe-15
Enter the term 1 coefficient value (x 10
-15
F). CC0; CC0?
TERM 2-C1
P/MXXX.XXe-27
Enter the term 2 coefficient value (x 10
-27
F/Hz). CC1; CC1?
TERM 3-C2
P/MXXX.XXe-36
Enter the term 3 coefficient value (x 10
-36
F/Hz
2
). CC2; CC2?
TERM 4-C3
P/MXXX.XXe-45
Enter the term 4 coefficient value (x 10
-45
F/Hz
3
). CC3; CC3?
ENTER THE
OFFSET LENGTH
None
OFFSET LENGTH
P/MXXX.XXX mm
Enter and display offset length of Open. COO; COO?
NEXT CAL STEP Changes the soft-key display to the CONFIRM
CALIBRATION PARAMETERS menu.
None
PORT X
SHORT DEVICE
ENTER THE
INDUCTANCE
COEFFICIENTS
None
TERM 1-
P/MXXX.XXe-12
Enter the term 1 coefficient value (x 10
-12
H). CC0; CC0?
TERM 2
P/MXXX.XXe - 24
Enter the term 2 coefficient value (x 10
-24
H/Hz). CC1; CC1?
TERM 3
P/MXXX.XXe-33
Enter the term 3 coefficient value (x 10
-33
H/Hz
2
). CC2; CC2?
TERM 4
P/MXXX.XXe-42
Enter the term 4 coefficient value (x 10
-42
H/Hz
3
). CC3; CC3?
ENTER THE
OFFSET LENGTH
None
OFFSET LENGTH
P/MXXX.XXX mm
Enter and display offset length of Short. COS; COS?
NEXT CAL STEP Changes the soft-key display to the CONFIRM
CALIBRATION PARAMETERS menu.
None
RECEIVER
CALIBRATION
(HARMONIC
APPLICATION)
CONNECT A THROUGH
LINE BETWEEN TEST
PORTS
None
RECEIVER PORT
PORT 1/ PORT 2/
PORT 3
Selects Receiver port to be calibrated. RCALRP1;
RCALRP2;
RCALRP3;
RCALRPX?
SOURCE PORT
PORT 1/ PORT 2/
PORT 3
Selects Source port to be calibrated. RCALSP1;
RCALSP2;
RCALSP3;
RCALSPX
BEGIN CAL Starts measurement and changes soft-key display to
the RECEIVER CALIBRATION 2 menu.
BEGR
RETURN Returns to the previous soft-key menu. None
CAL FRONT PANEL
KEY MENUS OPERATION
4-166 MS462XX OM