Specifications

Menu Name Function Description GPIB Mnemonic
SELECT
1 PATH 2 PORT
CALIBRATION
TYPE
FORWARD PATH
(S11, S21)
Corrects the forward transmission and reflection error
terms, ETF, EDF, ESF, ERF, (EXF) and changes
soft-key display to the SELECT USE OF ISOLATION
IN CALIBRATION menu.
C8T; CXX?
REVERSE PATH
(S22, S12)
Corrects the reverse transmission and reflection error
term, EDR, ESR, ERR, ETR, (EXR) and changes
soft-key display to the SELECT USE OF ISOLATION
IN CALIBRATION menu.
C8R; CXX?
SELECT
CALIBRATION
TYPE
FULL 12 TERM Corrects for all 12 error terms and changes soft-key
display to the SELECT USE OF ISOLATION IN
CALIBRATION menu.
C12; CXX?
1 PATH, 2 PORT Corrects for the 1 path, 2 port error terms and
changes soft-key display to the SELECT USE OF
ISOLATION IN CALIBRATION menu.
None
TRANSMISSION FREQ
RESPONSE
Corrects for the reflection only error terms and
changes soft-key display to the SELECT USE OF
ISOLATION IN CALIBRATION menu.
None
REFLECTION ONLY Corrects for the reflection-only error terms and
changes soft-key display to the SELECT USE OF
ISOLATION IN CALIBRATION menu.
None
SELECT
CALIBRATION
DATA POINTS
NORMAL (1601 POINTS
MAXIMUM)
Selects the standard calibration using a start-to-stop
frequency that provides for up to 1601 equally spaced
(except the last) points of data for the defined
frequency range. Changes the soft-key display to the
CAL FREQ RANGE menu.
NOC; SWP?
C.W. (1 POINT) Selects the single frequency (C.W.) calibration
sequence that provides for 1 data point at a selected
frequency. Changes the soft-key display to the
SINGLE POINT C.W. CALIBRATION menu.
CWC; SWP?
N-DISCRETE
FREQUENCIES (2 TO
1601 POINTS)
Selects the discrete frequency calibration mode that
allows inputting a list of from 2 to 1601 individual data
point frequencies. Changes the soft-key display to the
DISCRETE FILL menu.
DFC; SWP?
TIME DOMAIN
(HARMONIC)
Selects time domain harmonic frequency calibration
data points and changes the soft-key display to the
CALIBRATION RANGE menu.
TDC; SWP?
SEGMENTED SWEEP
(1601 POINTS
MAXIMUM)
Selects segmented sweep calibration data points and
changes the soft-key display to the CAL FREQ
RANGE 3 menu.
SSC; SWP?
CAL FRONT PANEL
KEY MENUS OPERATION
4-168 MS462XX OM