Specifications

4-14 AVG KEY MENUS This key provides entry to averaging and IF bandwidth functions.
Menu Name Function Description GPIB Mnemonic
AVERAGE 1
AVERAGE ON/OFF Toggles data averaging on or off. AON; AOF; AOF?
DATA AVERAGE:
XX MEA./POINT
Sets the number of measurements to be averaged. AVG; AVG?
TRACE SMOOTHING
ON/OFF
Toggles TRACE SMOOTHING on or off. SOF; SOF?
SMOOTHING 0-20%
OF SWEEP
Sets trace smoothing percentage. SON; SON?
NOISE FIGURE
AVERAGE ON/OFF
Toggles NOISE FIGURE AVERAGE on or off. NFAON; NFAOF;
NFAON?
NOISE FIGURE
AVERAGE XX
Sets the number of noise figure measurements to
be averaged.
NFAVEC; NFAVEC?
SELECT I.F.
BANDWIDTH
Changes the soft-key display to the I.F.
BANDWIDTH menu.
None
AVERAGE 2
Displays in the Frequency Translation Group Delay or the Transmission/Reflection Modes Only.
AVERAGE ON/OFF Toggles the averaging mode on or off. AON; AOF; AOF?
DATA AVERAGE
1 MEAS.
Enters the averaging count and turn it on. AVG; AVG?
AVERAGING TYPE
POINT-BY-POINT
SWEEP-BY-SWEEP
Selects the averaging type of Point-by-Point or
Sweep-by-Sweep.
SWAVG; PTAVG;
SWAVG?; AVGCNT?
RESET AVG. COUNT
(X SWEEP(S))
Resets the Sweep-by-Sweep averaging sweep
count.
RSTAVG
TRACE ON/OFF
SMOOTHING
Toggles trace smoothing on or off. SON; SOF; SOF?
SMOOTHING
X.XX% OF SWEEP
X POINTS
Displays the number of smoothing points. SON; SON?; SPTS?
SELECT I.F.
BANDWIDTH
Changes the soft-key display to the I.F.
BANDWIDTH menu.
None
AVG FRONT PANEL
KEY MENUS OPERATION
4-182 MS462XX OM