Specifications

ERROR AND STATUS ERROR
MESSAGES MESSAGES
MS462XX OM 5-7
Code Message Description Corrective Action
TOO HIGH TEMP CHANGE Temperature is too high to
change to
Decrease the need
temperature
H/W RETUNED SINCE CAL
BAD FILENAME File name is not acceptable Re-enter the file name
according to the rule
CAL MAY BE INVALID Calibration data may not be
applicable for the current
application
Check the needed cal
FLAT CAL MAY BE INVALID Flat calibration data invalid Redo a flat calibration
FLAT PWR TURNED OFF Flat power is off
HARD COPY OUTPUT ABORTED An abort of hardcopy out is
performed
CORRECTION UNAVAILABLE A corrective action is not
supported
TIME DOMAIN INVALID Time domain value is not valid
GATE MUST BE ON Gating must be on in Time
Domain for this function
SMOOTHING INVALID
MEMORY DATA INVALID Memory may be corrupted Contact manufacturer
NEED HARMONIC SWEEP RF Cal needs to be done in
Harmonic Sweep in order to be
applied in Time Domain
Perform RF Cal in
Harmonic Sweep
PARAMETER INVALID Input parameter is out of range Perform new cal
CORRUPT OR INCOMPATIBLE FILE File parameter data structure
is not recognized
Contact factory
NOT ALLOWED AT THIS TIME This function is not allowed
under the current setup
STORED SWP FREQS CLEARED The frequency list in Power
Sweep Gain Compression has
been cleared
HARD COPY DATA CAPTURE FAILED A hardcopy of data can't be
made
PRINTER BUSY Printer is busy in performing
other task
Check connection or wait
VALUE NOT FOUND The need value is not found in
the system
OUT OF SOURCE CONTROL RANGE The value is not in the control
range
Correct the value
ABORTED: INCOMPATIBLE FILES Not an acceptable file type
POWER OUT OF CALIBRATED RANGE Power is not in the range of
CAL
LOGO FILE NOT FOUND LOGO file is not on the disk Create a LOGO file and
reload
Table 5-1. General Error Messages (5 of 7)