Specifications

above (below) the marker. The delta symbol appears
above (below) the number.
Marker
Statistics
The marker statistics available in the marker menu
calculates and displays mean, peak to peak and
standard deviation on screen. Two mode selections
are available: one calculates the statistics between
the reference marker and the active marker
(Figure 6-14) and the other calculates them for the
entire display (Figure 6-15, on next page).
6-4 LIMITS Limit lines function as settable maximum and minimum indicators for
the value of displayed data. These lines are settable in the basic units
of the measurement on a channel-by-channel basis. If the display is
rescaled, the limit line(s) will move automatically and thereby
maintain their correct value(s).
Each channel has two limit lines (four for dual displays), each of which
may take on any value. Limit lines are either horizontal lines in
rectilinear displays or concentric circles around the origin in Smith
and polar displays.
DATA LIMITS
DISPLAYS
MS462XX OM 6-13
Figure 6-14. Marker Statistics, Between Reference Markers