The verification of the kit components is straight
forward. The components are first measured with
the MS462XX, then they are compared with the
data recorded on the disk. If the measured data
compares favorably with the recorded data (taking
tolerances into consideration), then the system is
known to be operating properly and providing accu
-
rate data.
There is one caution that you need to observe when
using Verification Kits. Because the verification
components have been characterized, you must han
-
dle them carefully so that you do not change their
known characteristics. Consequently, you should not
have them available for daily use. Rather, you
should only use them for the accuracy verification
checks taken every 6-to-12 months (or at any other
time the system’s integrity is in doubt).
This completes the discussion on calibration. Refer
to paragraph 7-4 for a procedure showing how to
calibrate the sliding load.