Specifications

Step 13. Loosen the FINE LOCK ring and turn the FINE
ADJ ring to position the gauge pointer 2-3 small di
-
visions on the “–” side of zero.
Step 14.
Turn the FINE LOCK ring clockwise to both tighten
the adjustment and place the pointer exactly to “0.
The Sliding Termination is now ready to use.
NOTES
· Ensure that the inner dial reads “0”
· The following step is not normally neces-
sary. It needs to be done only if the ad-
justment has changed since it was set at
the factory
Step 15.
With the 01-211 Flush Short installed, loosen the
COARSE LOCK and gently push the COARSE SET
adjustment rod in as far as it will go. This coarsely
sets the center conductor to be flush against the at
-
tached short. Return to Step 2.
Step 16.
The procedure for adjusting the male-connector
sliding termination is essentially the same as that
described above. The only difference is that you
must install the female adapter on the end of the
gauge shaft, over the center conductor. To install
this adapter, proceed as follows:
q
Zero-set the gauge as described in Step 2
q
Push the outer locking ring back toward the
gauge and turn it clockwise onto the exposed
threads
q
Loosen the lock ring one turn in a
counterclockwise direction
MEASUREMENT MEASUREMENT CALIBRATION
CALIBRATION SLIDING -TERMINATION
MS462XX OM 7-17