Specifications

tion frequency range. C.W.: Data is taken at one
point. N-DISCRETE FREQUENCIES (2 TO 1601
POINTS): This selection lets you specify a discrete
number of frequency points, from 2 to 1601. TIME
DOMAIN (HARMONIC): This selection is the cali
-
bration mode for low-pass time-domain or for Har
-
monic processing. It lets you select frequencies at in
-
teger (harmonic) multiples of the start frequency.
Step 8.
Press START, then enter 10 MHz, using either the
rotary knob or keypad and terminator keys.
Step 9.
Press STOP, then enter 3 GHz (MS4622X) or 6 GHz
(MS4623X), using either the rotary knob or keypad
and terminator keys.
The above 2 steps let you set your start and stop fre
-
quencies. For this example, press START, press 10 on
keypad, and hit the M/ms/cm terminator key. Perform
like operations for STOP, except make entry read 3
GHz or 6 GHz.
Step 10.
Press NEXT CAL STEP.
In this soft-key set, current settings for all parame-
ters appear in blue. If you do not need to make any
changes, you would press the START CAL soft key.
However, for this example, we will change them all.
Step 11. Press the following soft keys in sequence:
PORT 1 CONN
TYPE N (M)
PORT 2 CONN
TYPE N (F)
REFLECTION PAIRING
MIXED
Reflection Pairing lets you mix or match the Open
and Short reflection devices in the Calibration
Sequence menus. The MIXED choice lets you cali
-
brate using first an Open on one port and a Short on
the other, then a Short on one port and an Open on
the other. Conversely, MATCHED lets you calibrate
first using an Open on both ports then using a Short
on both ports.
STANDARD 2-PORT MEASUREMENT
(OSL) CALIBRATION CALIBRATION
7-20 MS462XX OM