Specifications

Step 12.
Press LOAD TYPE.
The next soft-key set lets you select either of two load
types, broadband or sliding. Broadband loads are
adequate for all but the most demanding reflection
measurements. They are easier to use and less expen
-
sive than sliding loads. If you choose a sliding load,
refer to paragraph 7-3 for a procedure on setting pin
depth. For this example, we will select BROADBAND
FIXED LOAD.
Step 13.
Press BROADBAND FIXED LOAD.
The next soft-key set, prompts you to enter impedance
and inductance values. For this example, we will use
the rotary knob to change the displayed value to 50
W. Alternatively, you can key in 50 Ohms. That is,
press 50 on the keypad and the X1 terminator key.
Step 14.
Press IMPEDANCE and set for 50W, using the ro-
tary knob.
Step 15.
Press NEXT CAL STEP.
Step 16.
Press THROUGH LINE PARAMETERS.
The next soft-key set lets you define the length of the
offset and the impedance of the throughline. For this
example, enter 0 mm for length and 50W for imped
-
ance.
Step 17.
Press OFFSET LENGTH and set for 0.0 mm, using
the rotary knob.
Step 18.
Press THROUGH LINE IMPEDANCE and set for
50 Ohms, using the rotary knob.
Step 19.
Press NEXT CAL STEP.
Step 20.
Press REFERENCE IMPEDANCE/REFERENCE
IMPEDANCE and set for 50W.
Step 21.
Press TEST SIGNALS.
MEASUREMENT STANDARD 2-PORT
CALIBRATION (OSL) CALIBRATION
MS462XX OM 7-21