Specifications

The next soft-key set lets you define the power level of
the signals at the two test ports.
(If you needed to calibrate the test port for power
flatness, you would press the FLATNESS CORREC
-
TION soft key to ON.)
Step 22.
Press RESUME CAL.
Step 23.
Press START CAL.
Step 24. Continue the calibration sequence by following the
prompts as they appear. Connect the appropriate
Isolation Devices, Broadband Loads, Opens, Shorts,
and Throughlines, when requested in the calibration
sequence.
STANDARD 2-PORT MEASUREMENT
(OSL) CALIBRATION CALIBRATION