Specifications

b.
Press DEVICE 2 LINE/MATCH.
Here you have another decision to make: whether
your calibration is to be LRL or LRM. The soft key
acts as a toggle.
c.
If you toggle such that LINE turns red, then en
-
ter the value for line 2.
This value depends on your frequency range.
d.
If you toggle MATCH red, observe that
FULLBAND appears.
This indicates that your reflective device covers the
full calibration range.
e. When you have made both selections, press
NEXT CAL STEP and go to Step 17.
Step 16.
If TWO BANDS was selected in Step 13, perform
the following:
a.
Press DEVICE 1 LINE 1 (REF) and enter the
value (typically 1.00 cm).
b.
Press DEVICE 2 LINE/MATCH.
Both here, and for the next choice, you have another
decision to make: whether your calibration is to be
LRL or LRM. For this selection, the soft key acts as a
toggle.
c.
If you toggle such that LINE turns red, then en
-
ter the value for line 2.
This value depends on your frequency range.
d.
If you toggle MATCH red, observe that
LOWBAND appears.
This indicates that your reflection device is a
low-band load. This load must have a passband
such that it passes all frequencies from the start to
the breakpoint (see below).
e.
Press DEVICE 3 LINE/MATCH. If device 3 is
a line, key in the value. If it is a match, the term
HIGHBAND will appear.
LRL/LRM MEASUREMENT
CALIBRATION CALIBRATION
7-30 MS462XX OM