Specifications

(If you needed to calibrate the test port for power
flatness, you would press the FLATNESS CORREC
-
TION soft key to ON.)
Step 22.
Press RESUME CAL.
Step 23.
Press START CAL.
Step 24. Continue the calibration sequence by following the
prompts as they appear. Mount the appropriate LRL
line substrates when requested in the calibration se
-
quence.
For the REFLECTIVE DEVICE and BROADBAND
LOAD prompts, remove all substrates from the UTF
and allow the lower jaws to short the center conduc
-
tor. Separate the connector blocks by at least an inch.
(The BROADBAND LOAD prompt only appears if
you selected to include isolation.)
Step 25. Store the calibration.
LRL/LRM MEASUREMENT
CALIBRATION CALIBRATION
7-32 MS462XX OM