Specifications

Chapter 2
Installation
2-1 INTRODUCTION This chapter provides information for the initial inspection and
preparation for use of the MS462XX Vector Network Measurement
System. Information for interfacing the MS462XX to the IEEE-488
General Purpose Interface Bus and reshipment and storage
information is also included.
2-2 INITIAL INSPECTION Inspect the shipping container for damage. If the container or
cushioning material is damaged, retain until the contents of the
shipment have been checked against the packing list and the
instrument has been checked for mechanical and electrical operation.
If the MS462XX is damaged mechanically, notify your local sales
representative or Anritsu Customer Service. If either the shipping
container is damaged or the cushioning material shows signs of stress,
notify the carrier as well as Anritsu. Keep the shipping materials for
the carrier’s inspection.
2-3 PREPARATION FOR USE No initial setup is required. After unpacking, the MS462XX is ready
for use. The MS462XX is equipped with automatic line-power sensing,
and will operate with any of the following line voltages: 100V, 120V,
220V, 240V +5%, –10%, 48–63 Hz, 350 VA. The MS462XX is intended
for Installation Category (Overvoltage Category) II.
MS462XX OM 2-3
Use two or more people to lift and move this equipment, or use an
equipment cart. There is a risk of back injury, if this equipment is
lifted by one person.
WARNING
When supplying power to this equipment, always use a three-wire
power cable connected to a three-wire power line outlet. If power is
supplied without grounding the equipment, there is a risk of re
-
ceiving a severe or fatal electric shock.
WARNING