Specifications

Step 13.
Ensure the AutoCal module is properly connected
between Ports 1 and 2, then press the START CAL
soft key (AutoCal main menu).
Follow the instructions and do not disturb the setup
during the calibration. A typical Full 2-Port Cal
with 401 data points should take about one minute.
Please note that you should not start a calibration
until the yellow LED on the AutoCal Module is lit.
This will ensure accurate calibration of the VNMS.
After the Calibration, select Save/Recall to save the
calibration data, as shown in the following steps. Al
-
ternatively, you could press Enter to apply the cali
-
bration.
Step 14. Press the Save/Recall key.
Step 15.
Press the SAVE\FRONT PANEL SETUP AND
CAL DATA ON HARD DISK (or ON FLOPPY
DISK) soft keys.
Step 16. Enter a file name using the name editor or an exter-
nal keyboard.