Specifications

Inappropriate setting of the Default Gateway IP Address will cause
the Scorpion system to appear to be locked up at start up. The
instrument will appear to stop working at the following message:
Application loaded successfully, starting system…
Once the Default Gateway IP Address has been inappropriately set,
the only way to restore the system is to use the debugger loader utility
program on the 2300-246 Boot Utility Diskette and the Hyper
Terminal Utility program (supplied with Microsoft Windows)
connected via the serial port on the rear panel. A null modem cable,
part number 800-441, is also required.
Procedures found in Chapter 3 of the MS462XX Programming Manual,
part number 10410-00204, guide you through the Ethernet setup and
configuration process.
2-7 EXTERNAL MONITOR
CONNECTOR
The rear panel VGA connector allows the internal display information
of the MS462XX to be connected to an external VGA monitor (either
color or monochrome). The pinout of this 15-pin Type D connector is
shown in Appendix B, Page B-10.
2-8 RACK MOUNT
INSTALLATION
To install the Option 1 Rack Mount rails, refer to the procedure below.
Step 1. Disconnect the line cord and any other attachments from the
instrument.
Step 2. Carefully place the instrument on a secure and stable work surface.
Step 3. Using a Phillips screwdriver, remove the two handles or four bumper
assemblies (and tilt bail, if installed) from the front of the unit, and
the four feet at the rear. Save the screws for later use.
NOTES
q
The green-headed screws are metric threads and must be used
only in the appropriately tapped holes
q
The feet, handles, and bumpers are not reused in this application
Step 4. Remove the side carrying handle screws and remove the handle.
Step 5. For Option 1A (System Console), install the two Rack Mount Handles
using the green-headed screws removed earlier (or use the ones from
PREPARATION FOR INSTALLATION
STORAGE AND/OR SHIPMENT