PARAMETERS TEST
This test verifies that each individual receiver channel operates
properly. Measurement calibration of the system is not required for
this test.
Preliminary Perform the steps in paragraph 10-4.
Test Setup Setup MS462XX as described below:
Step 1. Connect Ports 1 and 2 together using a high-quality
through line (top left).
Step 2.
Press Default (middle left) then CONTINUE to re
-
set the MS462XX to its factory-default settings.
Test
Procedure
Perform the test as described below:
Step 3. Observe the displayed FREQUENCY menu and
note that the START and STOP values are the low-
and high-end frequencies.
Step 4. Press Display (middle left).
Step 5.
Press DISPLAY MODE/DUAL CHANNEL1&3
(MS462XXA) or FOUR CHANNELS (MS462XXB).
Step 6.
Press RETURN.
Step 7.
Press GRAPH TYPE.
Step 8.
For MS462XA: Press MORE twice then LOG MAG
-
NITUDE, Ch 3 (bottom left), MORE, then LOG
MAGNITUDE.
For MS462XB: Press Ch 1, MORE twice then LOG
MAGNITUDE, Ch 2, MORE then LOG MAGNI
-
TUDE,Ch3,MORE then LOG MAGNITUDE, Ch
4, MORE twice then LOG MAGNITUDE.
Step 9.
Press RETURN.
Step 10.
Press SCALE/RESOLUTION
OPERATIONAL CHECKOUT MS462XA/B NON-RATIO
PROCEDURES PARAMETERS TEST
MS462XX OM 10-5
PORT 1 PORT 2
TEST PORT 1
TEST PORT 2
DataEntry
Power
Keyboard
Port 3
Channels
Enhancement
Stimulus
System
Port 1
Port 2
Probe Power
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch 1
Ch 2
Ch 4
Ch 3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save /
Recall
0
7
4
Display
Clr
Print
Soft
Keys
Default
DataEntry
Power
Keyboard
Port 3
Channels
Enhancement
Stimulus
System
Port 1
Port 2
Probe Power
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch 1
Ch 2
Ch 4
Ch 3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save /
Recall
0
7
4
Display
Clr
Print
Soft
Keys
Display
Data Entry
Power
Keyboard
Port 3
Channels
Enhancement
Stimulus
System
Port 1
Port 2
Probe Power
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch 1
Ch 2
Ch 4
Ch 3
Channels
Ch 1
Ch 2
Ch 4
Ch 3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save /
Recall
0
7
4
Display
Clr
Print
Soft
Keys