Specifications

Step 18. After at least one sweep has occurred, verify that
the maximum-value to minimum-value amplitude
slope (Figure 10-1, page 10-6) meets the specifica
-
tions shown below.
Model
High-End
Frequency
Test
Channel Slope
(b1/1 & b2/1)
Reference
Channel Slope
(a1/1 & a2/1)
MS4622A 3 GHz 5 dB 5 dB
MS4622B 3 GHz 5 dB 6 dB
MS4623A 6 GHz 7 dB 10 dB
MS4623B 6 GHz 7 dB 10 dB
Step 19. Verify that the minimum amplitude meets the speci
-
fications shown below.
NOTE
If desired, use the Readout Markers func-
tion (Marker key ( left) and Readout Marker
soft key) to obtain precise frequency and
amplitude values.
Model
Test Channel
(b1/1 & b2/1)
Reference Channel
(a1/1 & a2/1)
MS4622A -12 dB -11 dB
MS4622B -10 dB -13 dB
MS4623A -14 dB -11 dB
MS4623B -11 dB -14 dB
MS462XA/B NON-RATIO OPERATIONAL CHECKOUT
PARAMETERS TEST PROCEDURES
10-8 MS462XX OM