Specifications

Step 27. Enter the following formula in the cell that follows
the last data point (top left).
SORT(SUMSQ(X1:X401)/401)
Where X is the column letter in the spreadsheet
Step 28. Observe that the calculated result is as follows:
10 to 3000 MHz <0.008
3000 to 6000 MHz <0.018
Step 29. For MS4623A and MS4623B, perform steps 28 and
29; for MS4622A or MS4622B, skip to step 31.
Step 30. Press Freq key (bottom left).
Step 31.
Change STOP frequency to 6 GHz and START fre
-
quency to 3 GHz.
Step 32. Repeat steps 12 through 26.
OPERATIONAL CHECKOUT MS462XA/B HIGH
PROCEDURES LEVEL NOISE TEST
MS462XX OM 10-13
Data Entry
Power
Keyboard
Port 3
Channels
Enhancement
Stimulus
System
Port 1
Port 2
Probe Power
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch 1
Ch 2
Ch 4
Ch 3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save /
Recall
0
7
4
Display
Clr
Print
Soft
Keys
Freq