Specifications

10-7 MS462XC SYSTEM
DYNAMIC RANGE
TEST
This test verifies the System Dynamic Range of the test channels of
the MS462XC receiver.
Test Setup Set up the equipment as shown below and allow
them to warm up for at least 60 minutes.
Test
Procedure
Perform test as described below:
Step 1. Press Default key and then 0 key to reset the
MS462XC.
Step 2. Press Ch 3 key and then Display key.
Step 3. Select:
DISPLAY MODE
SINGLE CHANNEL
RETURN
GRAPH TYPE
LOG MAGNITUDE
Step 4. Press Config key.
Step 5. Select DATA POINTS and change the value to
51MAX PTS.
Step 6. Press Power key.
OPERATIONAL CHECKOUT MS462XC SYSTEM
PROCEDURES DYNAMIC RANGE TEST
MS462XX OM 10-17
DataEntry
Power
Keyboard
Port3
Channels
Enhancement
Stimulus
System
Port1
Port2
ProbePower
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch1
Ch2
Ch4
Ch3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save/
Recall
0
7
4
Display
Clr
Print
Noise
In
Noise
+28V
NOOPERATORSERVICE-
ABLEPARTSINSIDE.
REFERSERVICING TO
QUALIFIEDPERSONNEL.
WARNING
CAUTION
FORCONTINUED FIRE
PROTECTIONREPLACE
ONLYWITHSPECIFIED
TYPEAND RATEDFUSE.
IEEE488.2
GPIB
Dedicated
GPIB
SerialPort
Ethernet
VGA
SCSI-2
Ext
Analog
Output
Ext
Trigger
LPTLine Printer
ExtI/O
ExtAnalog
In
Ext
Source
Ext10 MHz
In
47-440Hz
85-264VAC
REPLACEFUSE ONLY WITH
SAMETYPE AND RATING
CAUTION
DONOT OPERATE
WITHPOWER CORD
UNGROUNDED
N274
A1
B1
A2 B2
MS46XC Rear Panel
MS462XC Front Panel
Anritsu Power Meter
Power Sensor
b1 b2
Coaxial Cable
Dedicated GPIB
GPIB Cable
Port 1
*Connect dashed lines as directed by procedure
43KC-10
Attenuator
*
Figure 10-4. Test Equipment Setup for System Dynamic Range Test