Specifications

Step 22. Allow one full sweep to occur, then press Hold key.
Step 23. Verify that the trace is below the REFERENCE
VALUE.
Step 24. Press Meas key and select USER DEFINED.
Step 25. Select:
CHANGE RATIO
b2 (Tb)
1 (UNITY)
Step 26. Remove the 43K-10 Attenuator with the coaxial ca
-
ble from b1 input and connect to b2 input.
Step 27. Repeat Step 17 to Step 23.
OPERATIONAL CHECKOUT MS462XC SYSTEM
PROCEDURES DYNAMIC RANGE TEST
MS462XX OM 10-19/10-20