Specifications

A-9 SEQUENCE EXECUTION There are two types of sequence execution: regular and auto. A
regular execution is initiated by pressing the Seq key and choosing
EXECUTE SEQUENCE. An auto execution of SEQUENCE 7 is
initiated automatically after power up, if SEQUENCE 7 exists and is
named “AUTO.”
Regular Execution
Step 1.
Press Seq/EXECUTE SEQUENCE/SEQUENCE1.
Auto Execution
Step 2. Press Seq then the following soft keys:
UTILITY
RENAME SEQUENCE
SEQUENCE 7
Step 3. Using the rotary knob, turn the knob and select the
characters “AUTO” and DONE; using an external
keyboard, type “Auto.
Step 4.
Reboot the system and SEQUENCE/AUTO will ex-
ecute.
Termination of a Sequence Execution
Step 5. Press the Clr/Local key to terminate the execution.
Execution Through GPIB
An execution of a sequence can be activated by GPIB mnemonics.
Please refer to the MS462XX Programming Manual for details.
SEQUENCE SEQUENCE
OPERATIONS EXECUTION
MS462XX OM A-9