Specifications

A-10 DISPLAY MENU
DURING A SEQUENCE
EXECUTION
This function turns on the display of menu transitions in the sequence
list, and also causes menu transitions to occur during sequence
execution. For example, the start frequency is set under the
FREQUENCY menu. When show menu is on, the name of the
frequency menu will be added to the content of the displayed sequence
list, and the menu will be displayed on the instrument during
sequence execution. For this example, Sequence 2 is new and will
contain start frequency 1.2 GHz, stop frequency 1.8 GHz, and display
phase for graph type. The creation steps are:
Step 1.
Press Seq/EDIT SEQUENCE/SEQUENCE 2.
Step 2.
Press Seq/CONTROL STATEMENT/SHOW
MENU ON.
Step 3.
Press Freq/START 1.2 GHZ/STOP 1.8 GHZ.
Step 4.
Press Display/GRAPH TYPE/PHASE.
Step 5. After creation, the sequence display is as shown at
left.
Step 6.
Press Seq/DONE SEQUENCE EDIT.
This sequence is stored in memory as
SEQUENCE 2/NAME 2.
DISPLAY MENU SEQUENCE
DURING SEQUENCE EXECUTION OPERATIONS
A-10 MS462XX OM
BEGIN OF SEQUENCE
SHOW MENU ON
FREQUENCY MENU
START 1.2 G/ns/m
STOP 1.8 G/ns/m
DISPLAY MENU
GRAPH TYPE MENU1
PHASE