Specifications

Menu Name Function Description GPIB Mnemonic
READOUT LIMIT
-GRAPH TYPE- Displays the currently selected graph type. None
UPPER LIMIT (REF)
XX.XX dB
Sets the UPPER LIMIT (REF) limit line. Changing
this value also moves the lower limit line by the
LIMIT DIFFERENCE amount.
LUP; LUP?
LOWER LIMIT (REF)
XX.XX dB
Sets the LOWER LIMIT dB limit line. Changing this
value also changes the LIMIT DIFFERENCE amount
relative to the UPPER LIMIT (REF) value.
LLO; LLO?
LIMIT DIFFERENCE
D (UPPER-LOWER)
XXX.XXXX dB
Sets the LIMIT DIFFERENCE amount. Changing
this value also changes the lower limit value relative
to the UPPER LIMIT (REF) value.
LFD; LFD?; LFD2;
LFD2?
INTERCEPTS AT
LOWER LIMIT:
XXX.XXXXXXXXXGHz:
XXX.XXXXXXXXXGHz:
XXX.XXXXXXXXXGHz:
XXX.XXXXXXXXXGHz
Displays at which frequencies the data intercepts the
lower limit. May be interpolated.
None
RETURN Returns to the previous soft-key menu. None
REFERENCE
PLANE Ch x
AUTO Automatically sets the reference delay such that the
cumulative phase shift is zero. This selection
unwinds the phase in a Smith chart display or
reduces the phase revolutions in a rectilinear display
to less than one.
RDA
DISTANCE
XX.XXX mm
Electrically repositions the measurement reference
plane, as displayed on the active channel, by a
distance value entered in millimeters. This selection
lets you compensate for the phase reversals
inherent in a length of transmission line connected
betwee
RDD
TIME
XX.XXX ps
Electrically repositions the measurement reference
plane by a distance value that corresponds to the
time in picoseconds.
RDT
SET DIELECTRIC
XXXXX
Changes the soft-key display to the SET
DIELECTRIC menu.
DIE
PHASE OFFSET
XXXXXX
Add a constant phase offset to the display data. POSET; POSET?
PER CHANNEL/PORT Changes the soft-key display to the REFERENCE
PLANE PER PORT menu.
None
RETURN Returns to the previous soft-key menu. None
DISPLAY FRONT PANEL
KEY MENUS OPERATION
4-32 MS462XX OM