Specifications

Menu Name Function Description GPIB Mnemonic
REFERENCE
PLANE PER
PORT
PORT X Select the port number. RPPORTNUM;
RPPORTNUM?
AUTO Changes the S-parameter being measured on the
active channel to the reflection S-parameter that
corresponds to the port number. Automatically sets
the reference delay such that the cumulative phase
shift is zero. This selection unwinds the phase in a
Smith chart display or reduces the phase revolutions
in a rectilinear display to less than one.
RDA
DISTANCE
XX.XXX mm
Electrically repositions the measurement reference
plane in all measurements as specified in PORT X
(above) by a distance value entered in millimeters.
This selection lets you compensate for the phase
reversals inherent in a length of transmission line
connected between.
RDD
TIME
XX.XXX ps
Electrically repositions the measurement reference
plane by a distance value that corresponds to the
time in picoseconds.
RDT
SET DIELECTRIC
XXXXX
Changes the soft-key display to the SET
DIELECTRIC menu.
DIE
PER CHANNEL/PORT Changes the soft-key display to the REFERENCE
PLANE CHx menu.
None
RETURN Returns to the previous soft-key menu. None
S-PARAMS
SETUP IN ALL
CHANNELS
SINGLE ENDED
S-PARAMS
Sets all the groups to SINGLE ENDED
S-parameters.
SETSE
BALANCED
DIFFERENTIAL
Sets all the groups to BALANCED DIFFERENTIAL
S-parameters.
SETBD
SINGLE ENDED/
BALANCED
DIFFERENTIAL
Sets all the groups to 3 PORTS MIXED MODE
S-parameters.
SETSB
HELP ON/OFF Displays an informational help screen. HELP1; HELP0;
HELPX?
RETURN Returns to the previous soft-key menu. None
SCALE GROUP
DELAY
RESOLUTION
XXX.XXX uS/DIV
Sets the resolution for the vertical axis of the active
channel displayed graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XXX fS
Sets the value at the reference line for the active
channel amplitude measurement on the displayed
graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE X Sets the reference line for the active channel
amplitude measurement. This is the line about which
the amplitude expands with different resolution
values.
REF; REF2; REF?;
REF2?
APERTURE X.X
PERCENT OF SWEEP
Sets the percent of frequency span over which group
delay is calculated.
APR; APR?
RETURN Returns to the previous soft-key menu. None
FRONT PANEL DISPLAY
OPERATION KEY MENUS
MS462XX OM 4-33