Specifications

Menu Name Function Description GPIB Mnemonic
SCALE LINEAR
MAG & PHASE
(CAL)
-LINEAR MAG-
None
RESOLUTION
XX.XXX U/DIV
Sets the resolution for the vertical axis of the active
channel displayed phase graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XXX PU
Sets the value at the reference line for the active
channel amplitude measurement on the displayed
graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel
amplitude measurement. This is the line about which
the amplitude expands with different resolution
values.
REF; REF2; REF?;
REF2?
-PHASE-
None
RESOLUTION
XX.XX DEG/DIV
Sets the resolution for the vertical axis of the active
channel displayed phase graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XX DEG
Sets the value by which the active channel phase
measurement is offset on the phase graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel phase
measurement on the phase graph. This is the line
about which the phase expands with different
resolution values.
REF; REF2; REF?;
REF2?
PHASE SHIFT
X.XX
Sets the value by which the active channel phase
measurement is shifted on the phase graph.
PHO; PHO?
RESUME CAL Resumes calibration. None
SCALE LINEAR
MAGNITUDE
RESOLUTION
XX.XXX U/DIV
Sets the resolution for the vertical axis of the active
channel displayed graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XX pU
Sets the value at the reference line for the active
channel amplitude measurement on the displayed
graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel
amplitude measurement. This is the line about which
the amplitude expands with different resolution
values.
REF; REF2; REF?;
REF2?
RETURN Returns to the previous soft-key menu. None
SCALE LINEAR
POLAR
RESOLUTION
XX.XXX U/DIV
Sets the resolution for the vertical axis of the active
channel displayed graph. Resolution can be set
incrementally using the keypad or rotary knob. The
center is fixed at 0 units; therefore, changing the
resolution also changes the reference value and
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XXX U
Sets the value at the reference line for the active
channel amplitude measurement on the displayed
graph.
OFF; OFF2; OFF?;
OFF2?
FIXED REFERENCE
LINE
None
SELECT POLAR
CHART MODE
XXXXXXXX
Changes the soft-key display to the SELECT
POLAR CHART MODE menu.
PCP
RETURN Returns to the previous soft-key menu. None
FRONT PANEL DISPLAY
OPERATION KEY MENUS
MS462XX OM 4-35