Specifications

Menu Name Function Description GPIB Mnemonic
SCALE LOG MAG
& PHASE
-LOG MAG-
None
RESOLUTION
XX.XX dB/DIV
Sets the resolution for the vertical axis of the active
channel displayed graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XX.XX dBc/dBm
Sets the value at the reference line for the active
channel amplitude measurement on the
log-magnitude graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel
amplitude measurement on the log-magnitude
graph. This is the line about which the amplitude
expands with different resolution values.
REF; REF2; REF?;
REF2?
-PHASE-
None
RESOLUTION
XX.XX DEG/DIV
Sets the resolution for the vertical axis of the active
channel displayed phase graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XX DEG
Sets the value at the reference line for the active
channel amplitude measurement on the phase
graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel phase
measurement on the phase graph. This is the line
about which the phase expands with different
resolution values.
REF; REF2; REF?;
REF2?
PHASE SHIFT
X.XX DEG
Sets the value by which the active channel phase
measurement is shifted on the phase graph.
PHO; PHO?
RETURN Returns to the previous soft-key menu. None
SCALE LOG MAG
& PHASE(CAL)
-LOG MAG-
None
RESOLUTION
XX.XX dB/DIV
Sets the resolution for the vertical axis of the active
channel displayed graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XX.XX dBc/dBm
Sets the value at the reference line for the active
channel amplitude measurement on the
log-magnitude graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel
amplitude measurement on the log-magnitude
graph. This is the line about which the amplitude
expands with different resolution values.
REF; REF2; REF?;
REF2?
-PHASE-
None
RESOLUTION
XX.XX DEG/DIV
Sets the resolution for the vertical axis of the active
channel displayed phase graph.
SCL; SCL2; SCL?;
SCL2?
REFERENCE VALUE
XXX.XX DEG
Sets the value at the reference line for the active
channel amplitude measurement on the phase
graph.
OFF; OFF2; OFF?;
OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel phase
measurement on the phase graph. This is the line
about which the phase expands with different
resolution values.
REF; REF2; REF?;
REF2?
PHASE SHIFT
X.XX DEG
Sets the value by which the active channel phase
measurement is shifted on the phase graph.
PHO; PHO?
RESUME CAL Resumes the calibration. None
DISPLAY FRONT PANEL
KEY MENUS OPERATION
4-36 MS462XX OM