Specifications

Menu Name Function Description GPIB Mnemonic
SET DIELECTRIC
AIR
(1.000649)
Calculates reference delay based on the dielectric
constant of air (1.000649).
DIA
POLYETHYLENE
(2.26)
Calculates reference delay based on the dielectric
constant of polyethylene (2.26).
DIP
TEFLON
(2.10)
Calculates reference delay based on the dielectric
constant of teflon (2.1).
DIT
MICROPOROUS
TEFLON (1.69)
Calculates reference delay based on the dielectric
constant of microporous Teflon (1.69).
DIM
OTHER
XXXX.XX
Calculates reference delay based on the value you
enter.
DIE
RETURN Returns to the previous soft-key menu. None
SET SEGMENT
OFFSETS
-GRAPH TYPE- None
HORIZONTAL
XXX.XXXXXXX GHz
Sets the horizontal segment offset. SLH; SLH?
VERTICAL
XX.XXXXXXXX dB
Sets the vertical segment offset. SLV; SLV?
SINGLE LIMIT
-GRAPH TYPE- None
TEST LIMITS Changes the soft-key display to the TEST LIMITS
menu.
None
SEGMENTED LIMITS Changes the soft-key display to the SEGMENTED
LIMITS menu.
None
MORE Changes the soft-key display to the appropriate
SINGLE LIMITS menu.
None
RETURN Returns to the previous soft-key menu. None
SINGLE LIMITS
GROUP DELAY
-GROUP DELAY- None
UPPER LIMIT
XX.XXXXXX s
ON/OFF
Turns the Upper Limit line on or off for the active
channel, and sets its value. For your convenience,
the arbitrarily set limit lines allow you to delineate a
go/no go line on your Group Delay display beyond
which the measured values are unacceptable.
UPL1; UPL0
LOWER LIMIT
XX.XXXXXX s
ON/OFF
Turns the Lower Limit line on or off for the active
channel, and sets its value.
LOL1; LOL0
READOUT LIMIT Changes the soft-key display to the READOUT
LIMIT menu.
None
DISPLAY LIMITS
ON/OFF
Enables both previously set limit lines to appear for
the active channel.
LON; LOF; LON?
TEST LIMITS Changes the soft-key display to the TEST LIMITS
menu.
LTST
SEGMENTED LIMITS Changes the soft-key display to the SEGMENTED
LIMITS menu.
None
RETURN Returns to the previous soft-key menu. None
DISPLAY FRONT PANEL
KEY MENUS OPERATION
4-42 MS462XX OM