Specifications

4 Product Brochure
l
MX3701xxA
Additive White Gaussian Noise (AWGN) IQproducer
This GUI-based application software is used to generate AWGN
waveform pattern files optimized for each communication
system for the Dynamic Range Test, etc.
The AWGN waveform pattern file is created by setting the same
bandwidth and sampling rate as the combined waveform pattern
(Wanted Signal) and a multiplier of the Wanted Signal.
Specifying the combined waveform pattern (Wanted Signal) from
the waveform pattern for the desired communication method
automatically sets the Wanted Signal bandwidth and sampling rate.
The resultant AWGN waveform pattern and an existing waveform
pattern can be combined, which is useful for measuring base-
station dynamic range.
<Configurable Parameters>
(With Specified Wanted Signal)
AWGN BW/Wanted Signal BW ratio
(With Unspecified Wanted Signal)
Wanted Signal BW,
AWGN BW,
Sampling Rate
Main Parameter Settings
(1) Wanted Signal BW: Wanted Signal bandwidth
Setting range: 0.0010 to 120.0000 MHz
(2) AWGN BW (B)/Wanted Signal BW (A):
Magnification of AWGN to Wanted Signal
Setting range: 1.0, 1.5, 2.0, 2.5
(3) Sampling Rate:
Setting range: 0.0200 to 160.0000 MHz
Becomes same value as Wanted Signal
(4) AWGN BW (B): Bandwidth of AWGN
Calculated automatically from (1) and (2) with following
limitation:
Limit range: 0.001 to 20.000 MHz and Sampling rate/2 max.,
20.001 to 120.000 MHz and Sampling rate max.
IQproducer Main Screen
AWGN IQproducer Setting Screen
Standard accessory
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com