Network Router User Manual

8
Low Port Power
3652A Calibration Kit (K-Connector)
Corrected System Performance and Uncertainties:
MS202x/3xC Model with 12-term SOLT calibration including
isolation using 3652A Calibration Kit
Frequency Range (GHz) Directivity (dB)
≤ 5 > 42
≤ 15 > 36
≤ 20* > 32
Frequency Range (GHz) Typical Low Port Power (dBm)
≤ 3 –25
≤ 6 –25
≤ 20 –25
Measurement Uncertainties
The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration.
Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range,
and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination
were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.
0.1
1
10
100
S11 Magnitude
S11 Phase
S21 Magnitude
S21 Phase
0.1
1
10
-40 -35 -30 -25 -20 -15 -10 -5 0
0.1
10
100
-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0
0.01
0.1
1
1
10
-80 -70 -60 -50 -40 -30 -20 -10 0
Uncertainty 0.5 MHz to 20 MHz
Uncertainty 20 MHz to 3 GHz
Uncertainty 3 GHz to 6 GHz
Uncertainty 6 GHz to 20 GHz
Uncertainty 0.5 MHz to 20 MHz
Uncertainty 20 MHz to 3 GHz
Uncertainty 3 GHz to 6 GHz
Uncertainty 6 GHz to 20 GHz
Uncertainty 0.5 MHz to 20 MHz
Uncertainty 20 MHz to 3 GHz
Uncertainty 3 GHz to 6 GHz
Uncertainty 6 GHz to 20 GHz
Uncertainty 0.5 MHz to 20 MHz
Uncertainty 20 MHz to 3 GHz
Uncertainty 3 GHz to 6 GHz
Uncertainty 6 GHz to 20 GHz
* N Connector guaranteed to 18 GHz, typical > 18 GHz