User`s guide
MAINTENANCE MANUAL
Site Master™ Models S113C, S331C and S115BQ
Antenna and Cable Analyzers
1. INTRODUCTION
This manual provides maintenance instructions for
the Site Master S113C/S331C and S115BQ Antenna
and Cable Analyzer. It describes the product and
provides performance verification procedures, parts
replacement procedures, and a replaceable parts list.
2. DESCRIPTION
The Site Master (Figure 1) is a hand held SWR/RL
(standing wave ratio/return loss), and Dis
-
tance-To-Fault measurement instrument. It com
-
bines a synthesized source, VSWR Bridge, and re
-
ceiver on a single printed circuit board (PCB). An
optional power monitor is also available. A block dia
-
gram is shown in Figure 2.
3. PERFORMANCE VERIFICATION
Paragraphs 4 through 9 contain tests that can be
used to verify the performance of the Site Master
models S113C/ S331C and S115BQ having any ver
-
sion of firmware.
3.1. Initial Setup for Testing
1. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
2. Release the
ESCAPE/CLEAR key and use
the Up/Down arrow key to adjust the
contrast to give a readable display.
Figure 1. Site Master Model S331C
490 JARVIS DRIVE · MORGAN HILL, CA 95037-2809
P/N: 10580-00062
REVISION A
PRINTED: FEBRUARY 2002
COPYRIGHT 2002 ANRITSU CO.