Stomach Capsule User Manual Page | 115
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
Message
Solution
information
9.1.1.1.10. "Print
error"
prompt
box
Check printer power; check printer settings
Warning
If the solutions provided above do not address the problem, contact AnX
Robotica Corp. for assistance.
9.2 Problems with Translational Rotation Platform
Problem
Cause
Action
The power switch is on but
the system is not powered
Power cord is loosened
Check if the power cord is
properly connected
Fuse is burnt
Check and replace the fuse,
if necessary
Equipment failure
Contact technical support
personnel
The video monitoring
camera is lost, no image
found
USB connection cable is
loosened
Check the USB connection
cable of the camera
USB extension cable is
damaged
Replace USB extension
cable or contact technical
support personnel
Equipment failure
Contact technical support
personnel
The linear motion of three
axes is accompanied with
noise or jamming
Equipment is overloaded
Reset the related equipment
driver
Equipment failure
Contact technical support
personnel
Equipment emergency stop
of three-axis linear motion
Equipment error
Reset the related equipment
driver
Equipment failure
Contact technical support
personnel
Large difference of
three-axis motion positions
Equipment error
Reset the related equipment
driver
Equipment failure
Contact technical support
personnel
Over-speed of three-axis
Equipment error
Reset the related equipment