User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 116
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
Problem
Cause
Action
linear motion
driver
Equipment failure
Contact technical support
personnel
Driver inhibition error
Equipment error
Reset the related equipment
driver
Equipment failure
Contact technical support
personnel
9.3 Problems with Capsule
Problem
Cause
Action
When activating
capsule, LED light
does not lit up
Battery has no power or
capsule damaged
Replace the capsule
Warning: replacement of
the battery is prohibited.
Capsule is not activated
Activate the capsule again
with infrared light
Equipment failure
Contact technical support
personnel
Real-time image is not
transmitted to
computer
Capsule is not in working
state (all twelve cycles on the
monitor are hollow rather
than solid)
Issue operation commands
through computer
Problems with Data Recorder
See below actions
9.4 Problems with Data Recorder
Problem
Cause
Action
The power switch is on
but LED and monitor
do not lit up
The battery has no power
Recharge the recorder for 8
hours and use another
recorder for the procedure.
Warning: replacement of
the battery is prohibited.
Equipment failure
Contact technical support
personnel
Real-time image is not
transmitted to
computer
Capsule is not activated (the
circle in the upper left corner
of the monitor is hollow
rather than solid)
Activate the capsule with
infrared light
Capsule is not in working
state (all twelve cycles on the
monitor are hollow rather
Issue related operation
commands through the
computer