Operating instructions
INTRODUCTION 
20 Sep 12  1-5 
DE-7200VFDCentrifuge 
the upper setpoint, the compressor and condenser motors are cycled on to provide cooling. Upon 
reaching the low temperature setting, the air conditioner is shut down. The  evaporator fan 
operates continuously whenever  the cabinet is energized to maintain a continuous flow  of 
conditioned air over cabinet components. 
Purge System (Hazardous Environment Only) 
For operation in a hazardous environment, the control cabinet is equipped with a purge system. 
The purge system ensures that the cabinet interior remains free of potentially hazardous gases by 
filling the cabinet with filtered air and maintaining positive pressure to exclude ambient gases. 
During startup, before power is permitted to enter the cabinet, the cabinet purge system initiates a 
16-minute rapid purge cycle at 16CFM. When the rapid purge  cycle is completed, the purge 
system automatically closes a switch that supplies power to the cabinet. 
Following the rapid purge, the purge system maintains a positive pressure inside the cabinet to 
exclude  any potentially hazardous  gases. Minimum pressure and purge flow sensors will shut 
down power if the purge system is not maintaining satisfactory pressure in the cabinet. For further 
details of purge system operation, refer to Section 4. 
Vibration Switch 
The vibration switch (Figure 1-4) is a safety device designed to protect personnel and equipment 
by shutting down the centrifuge in case of excessive vibration. Normally, the switch contacts are 
held closed by a magnetic latch. However, strong vibration or a shock of 2 Gs will overcome the 
magnetic latch, causing the switch armature to break away from the normally closed position, 
providing an input to the PLC. A reset button on the side of the switch must then be manually 
pressed to close the contacts and re-engage the magnetic latch. 
WARNING! 
EXCEEDING THE 2 G SETTING CAN CAUSE INJURY TO 
PERSONNEL OR DAMAGE TO EQUIPMENT. 
DO NOT RAISE THE SETTING. 
The vibration switch is mounted on the centrifuge base in an orientation that is most affected by 
out-of-balance vibration of the bowl assembly. Clogging of the conveyor or worn bearings may 
produce sufficiently high vibration to trip the switch. 
Figure 1-4 Vibration Switch 
Case and Base 
When the top cover is closed, the case provides a sealed, protective enclosure that fully 
surrounds the bowl assembly. The liquid discharge chute and solid discharge chute are installed 
at the bottom of the lower case half. Mating baffles installed inside the top and bottom case halves 
separate the solid and liquid phases. Bolts secure the top and bottom halves together; a gasket in 
the bottom case half seals the two halves. 










