Operating instructions
MAINTENANCE 
31 Dec 13  5-7 
DE-7200VFDCentrifuge 
PURGE SYSTEM 
Bypassing Purge System 
The purge unit may be bypassed when necessary to apply electric power to the centrifuge while 
the control cabinet door is open, but this should be done only after verifying  that the 
centrifuge is in a non-hazardous area. After completing the work, the purge system must be 
restored to operation and satisfactory operation confirmed. Refer to Section 4 for the purge 
system bypass procedure. 
DANGER! HIGH VOLTAGE! USE EXTREME CAUTION WHEN OPERATING 
CENTRIFUGE WITH PURGE SYSTEM BYPASSED. DANGEROUSLY HIGH 
VOLTAGE WILL BE PRESENT IN CONTROL CABINET IF DOOR IS OPENED 
WHILE POWER IS APPLIED. 
EXPLOSION HAZARD! BE CERTAIN THAT SURROUNDING ATMOSPHERE IS 
CLEAR OF ALL POTENTIALLY EXPLOSIVE GASES BEFORE OPENING 
CONTROL CABINET DOOR. 
SAFETY GLASSES MUST BE WORN AT ALL TIMES WHILE PERFORMING 
ANY MAINTENANCE PROCEDURE. FAILURE TO WEAR SAFETY GLASSES 
MAY RESULT IN SERIOUS EYE INJURY OR PERMANENT LOSS OF VISION. 
Inspection and Test Procedures 
Periodic inspection and test procedures, supplemented by any additional requirements 
imposed by local codes, are recommended. The following tests should be performed at 
least every 6 to 24 months depending on site conditions. 
Visual Inspection and Checks 
1.  Inspect condition of relief valve and spark arrestor. Remove all debris and corrosion. 
2.  Check/drain air supply filters on cabinet exterior. 