TROUBLESHOOTING Document No. DER05105 
Effective Date  15 Jul 08  Page 7 of 16 
Alarm and Fault Messages 
Cause  Corrective Action 
Communication Failure with Solid End Bearing Temp Sensor (Cont’d) 
Poor connection at liquid end 
transmitter/intrinsic barrier or defective 
transmitter 
If reading is constantly 392°F (200°C), check and 
correct connection; replace transmitter, if defective. 
Defective bowl VFD  Check and replace VFD, if defective. 
Control Program Downloaded from Memory Card to Controller 
Re-loading of PLC program from 
memory card completed 
Start centrifuge, and resume operation. 
Controller Internal Battery Low, Replace 
PLC battery connections not secure  Check connections, and correct, if necessary. 
PLC battery low  Replace battery. 
Conveyor High Torque Alarm, Reduce Feed Rate 
Conveyor torque exceeds pre-set limit  Reduce feed rate. 
Conveyor High Torque Fault and Shutdown, Perform Cleanout and Reduce Feed Rate 
Conveyor torque exceeds pre-set limit  Following automatic shutdown, perform cleanout 
procedure to remove solids buildup, and re-start 
centrifuge at reduced feed rate. 
Conveyor VFD Alarm, Ref Troubleshooting Section of User Manual 
Temperature sensor signal cable 
disconnected 
Check and correct connection (refer to DER03022 
and electrical schematic diagram in Section 11). 
VFD temperature too high  Check Conveyor VFD status screen for temperature 
and alarm; shut down, lock out, and tag out 
centrifuge to determine and correct cause of 
overheating. 
VFD load too high  Reduce pump feed rate. 
Conveyor VFD Communications Error 
Confirm that green LED next to Ethernet cable is 
lighted, indicating that hub is functional. 
Check connections between conveyor VFD and 
PLC, and correct any defects. 
Conveyor VFD not communicating with 
PLC 
Replace conveyor VFD, if no other defect is found. 
Conveyor VFD Fault and Shutdown, Ref Troubleshooting Section of User Manual 
VFD fault detected, causing shutdown  Check and replace conveyor VFD. 