Specifications

INSTALLATION INSTRUCTIONS Document No. DER03022
Effective Date 23 Dec 08 Page 5 of 15
SHIPPING COMPONENTS
Note! Do not discard the rubber strips between bearing pillow blocks, cover,
and base after removal. These strips must be re-installed whenever the
machine is moved to prevent damage to the base and rotating assembly
bearings during transit.
Following final positioning and leveling of the centrifuge, soft rubber strips inserted between the
bearing pillow blocks and the base and hard rubber strips between the case cover and base must
be removed. The strips prevent damage to the rotating assembly bearings during transit.
To remove the rubber strips, proceed as follows:
1. Remove all bolts securing top cover along both sides, but do not remove bolts on both ends of
cover.
2. Remove bolts securing bearing pillow blocks to base and bolts securing upper halves of pillow
blocks to lower halves (Figure 3022-4).
3. Install two jack bolts (Figure 3022-5) in top half of each pillow block, and thread bolts through
lower halves.
Figure 3022-4 Pillow Block and Case Cover Attachment