Specifications
  Document No.  DER05025 
  Effective Date  28 Mar 08 
  Supersedes
  30 Aug 07 
  Page No.  1 of 4 
CONVEYOR BEARING REPLACEMENT 
DE-7200 CENTRIFUGE 
GENERAL 
This procedure describes replacement of the conveyor bearing. This procedure requires removal 
of the conveyor assembly from the bowl assembly. Excessive noise or vibration during centrifuge 
operation may indicate a defective bearing. If such conditions are evident, the conveyor should be 
removed and the bearing inspected for looseness and replaced if necessary. 
Bearing replacement must be performed by properly trained and qualified personnel. To prevent 
contamination of the new bearing and internal components of the centrifuge, the replacement 
procedure must be performed in a clean environment. 
Note! All attaching hardware must be correctly tightened. Refer to the torque 
chart in 
DER05029 for hardware tightening requirements. 
REMOVAL 
The liquid end of the conveyor is supported by a roller bearing fitted within a housing in the 
conveyor. The solid end is supported by the conveyor drive gearbox bearing. If the liquid end 
bearing is found defective, always check for gearbox shaft looseness. The gearbox shaft bearing 
must also be replaced if shaft looseness is found. 
To remove the conveyor bearing, proceed as follows: 
WARNING! TO AVOID SERIOUS PERSONAL INJURY BE SURE EQUIPMENT 
IS LOCKED OUT, TAGGED OUT, DE-ENERGIZED, AND HAS STOPPED 
ROTATING BEFORE PERFORMING MAINTENANCE AND/OR ADJUSTMENTS. 
1.  Shut down, lock out, and tag out electric power to the centrifuge (refer to DER04098). Allow 
rotating assembly to coast to a full stop. 
2.  Disconnect feed piping from feed component. 
3.  Remove screws securing belt guard to base, and lift and remove guard. 
4.  Remove screws securing case cover to case, and open case cover. 
5.  Remove conveyor drive gearbox in accordance with the gearbox removal procedures in 
DER05037. 
6. Hoist liquid end of rotating assembly until rotating assembly is vertical (Figure 5025-1), and 
then place solid end on a stable flat surface. Maintain suitable support of rotating assembly in 
vertical position throughout remainder of conveyor removal procedure. 