Installing the Test Set
Step 2. Meet Electrical and Environmental
Requirements
1. The line power module on your test set is an autoranging input. It is
designed to be used with an ac power source with a nominal voltage
of either 115 V or
230 V.
2. Ensure that the available ac power source meets the following
requirements:
90 to 250 Vac, 48 to 66 Hz, 40 W
CAUTION This product has an autoranging line voltage input. Be sure the supply
voltage is within the specified range.
If the ac line voltage is not within these ranges, use an
autotransformer that provides third-wire continuity to
earth-ground.
3. Ensure that the operating environment meets the following safety
requirements for the following conditions:
• indoor use
• altitude up to 15,000 feet (4,572 meters)
• temperature range of 0° C to 55° C
• maximum relative humidity 80% for temperatures up to 31° C,
decreasing linearly to 50% relative humidity
• enclosure protection, IP 20, according to IEC 529
WARNING This product is designed for use in INSTALLATION CATEGORY
II, and POLLUTION DEGREE 2, per IEC 101 and 664