User Manual

- 7 -
Short Press
or Long
Press
In manual temperature compensation (MTC) mode: Short press to adjust
temperature, long press to adjust swiftly.
In parameter setting mode: press to change the numbering of parameters
in main menu and sub-menu.
In sub-menu, press to change parameters and settings.
Connectors
The meter adopts 8-pin connector, into which pH, ORP, and conductivity electrode can be connected. When
connected, the meter will automatically switch to the correspondent measurement mode.
Display Mode
3.4.1 Stable Reading Display Mode
When the measuring value is stable, the screen displays as shown in
Figure-2. If does not appear or is flashing, that means the measuring
value has not been stable. Users should wait for the smiley face and not
record the readings or conduct calibrations at that moment.
3.4.2 Auto-Lock Display Mode
In parameter setting P4.6, select “On” to turn on the auto-lock display mode,
In which the reading will be automatically locked after the measuring value
has been stable for more than 10 seconds, and the HOLD icon will come up
as shown in Figure-3. Short press to cancel the hold.
4 pH MEASUREMENT
Information regarding the pH Electrode
The instrument is equipped with a 201T-S 3-in-1 Combination pH Electrode. Its built-in temperature sensor allows
for auto. temperature compensation. There is a vial stored with 3M KCL storage solution on top of the electrode.
Before using, loosen the cap of the vial, take out the electrode and rinse it in distilled water and then do measurement.
When not in use, place the electrode back to the vial and tighten the cap so that the pH sensor can be stored in the
best condition. When the pH electrode is dipped into the test sample solution, stir it for a few seconds to remove
potential air bubbles inside the probe to help the measurement get stabilized quickly.
When connecting the electrode into the instrument, please slowly rotate the connector, identify the location of the
mount to insert it, and then screw it on. Please note that do not pull on the cables in case of poor contact.
Please keep the connector clean and dry. For detailed information of the electrode’s maintenance, please refer to
section 4.5.
Information regarding pH Calibration
4.2.1 Standard Buffer Solutions
The instrument adopts two series of standard buffer solutions: USA and NIST as shown in Table-2. Users can select
which one to use in P1.1 (refer to section 7.3).
Figure-3
Figure-2