Gas Absolute Pressure: The M Series ow meters utilize an absolute pressure sensor to measure 
the line pressure of the gas ow being monitored. This sensor references hard vacuum and accurately 
reads line pressure both above and below local atmospheric pressure. This parameter is located in 
the upper left corner of the display under the dynamic label “PSIA”. This parameter can be moved to 
the primary display by pushing the button just above the dynamic label (top left). The engineering unit 
associated with absolute pressure is pounds per square inch absolute (PSIA). This can be converted 
to gage pressure (psig = the reading obtained by a pressure gauge that reads zero at atmospheric 
pressure) by simply subtracting local atmospheric pressure from the absolute pressure reading:
      PSIG = PSIA – (Local Atmospheric Pressure)
The ow meters use the absolute pressure of the gas in the calculation of the mass ow rate. For 
working in metric units, note that 1 PSI = 6.89 kPa.
Gas Temperature: The M Series ow meters also utilize a temperature sensor to measure the line 
temperature  of  the  gas  ow  being  monitored. The  temperature  is  displayed  in  engineering  units  of 
degrees Celsius (°C). The ow meters use the temperature of the gas in the calculation of the mass 
ow rate. This parameter is located in the upper middle portion of the display under “°C”. This parameter 
can be moved to the primary display by pushing the top center button above “°C”.
Volumetric Flow Rate: The volumetric ow rate is determined using the Flow Measurement Operating 
Principle described elsewhere in this manual. This parameter is located in the lower left corner of the 
display over “Volume”. This parameter can be moved to the primary display by pushing the “Volume” 
button (lower left). In order to get an accurate volumetric ow rate, the gas being measured must be 
selected (see Gas Select Mode). This is important because the device calculates the ow rate based on 
the viscosity of the gas at the measured temperature. If the gas being measured is not what is selected, 
an incorrect value for the viscosity of the gas will be used in the calculation of ow, and the resulting 
output will be inaccurate in direct proportion to the ratio between the two gases viscosities.
Mass Flow Rate: The mass ow rate is the volumetric ow rate corrected to a standard temperature 
and pressure (typically 14.696 psia and 25°C). This parameter is located in the lower middle portion 
of the display over “Mass”. This parameter can be moved to the primary display by pushing the button 
located below “Mass” (bottom center). The meter uses the measured temperature and the measured 
absolute pressure to calculate what the ow rate would be if the gas pressure was at 1 atmosphere and 
the gas temperature was 25°C. This allows a solid reference point for comparing one ow to another.
Flashing Error Message: Our ow meters and controllers display an error message (MOV = mass 
overrange, VOV = volumetric overrange, POV = pressure overrange, TOV = temperature overrange) 
when a measured parameter exceeds the range of the sensors in the device. When any item ashes 
on the display, neither the ashing parameter nor the mass ow measurement is accurate. Reducing 
the value of the ashing parameter to within specied limits will return the unit to normal operation and 
accuracy.