Volume Units (0 to 6) (40403)
Determine the unit of measure for the volumetric Applica! on Types.
1 = Cubic Feet
4 = Cubic Meters 7 = Barrels
2 = Million Cubic Feet 5 = Liters
3 = Gallons 6 = Cubic Inches
Decimal Place (0 to 3) (40404)
Used to set the number of decimal places included in the Calculated Reading.
The Calculated Reading will always be returned as a whole number.
Example: a Calculated Reading of 1126.658 (gallons, "
3
, etc.) will be returned
as follows:
Decimal Place = 0 Volume = 1127 (rounded to nearest whole number)
Decimal Place = 1 Volume = 11267 (divide by 10 to get true result)
Decimal Place = 2 Volume = 112666 (divide by 100 to get true result)
Decimal Place = 3 Volume = 1126658 (divide by 1000 to get true result)
Maximum Distance (0 to 10364 mm) (40405)
Sets the distance (beginning from the sensor face) to the point where the sen-
sor will stop looking for target signals. Targets detected beyond the Maximum
Distance value will be ignored by the sensor.
Full Distance (0 to 10364 mm) (40406)
Sets the distance (beginning from the sensor face) to the point where the tank
is considered full.
Empty Distance (0 to 10364 mm) (40407)
Sets the distance (beginning from the sensor face) to the point where tank is
considered empty. This will typically be the same as the tank depth unless the
sensor is mounted on a stand pipe.