- Acquisition Mode: this is a drop-down menu to select the measure type. Available options
are: 1D (the system returns just one level measure), 2D (the system returns level measures
on a 1D grid) or 3D (the system returns level measures on a 2D grid).
- Acquisition Speed: this is a drop-down menu allowing the selection of the rate of the radar
acquisition and hence, the frequency of level meter results updates. Three speed levels are
available: LOW (acquisition rate approximately of 13 Hz), MEDIUM (acquisition rate
approximately of 103 Hz) and HIGH (acquisition rate approximately of 188 Hz).
- Min. Range / Max. Range / Min. Phi / Max. Phi / Min. Theta / Max. Theta: these are minimum
and maximum values of distance, azimuth angle and elevation angle within which the level
survey must be limited. If 1D Acquisition Mode has been selected, only the minimum and
maximum range are used by the application; if 2D acquisition is set, both range and theta
limits are used. Finally, all the limits are necessary if the 3D acquisition mode has been
selected.
By clicking the Configure button, the acquisition is configured and the Acquisition State of the
application goes from READY to CONFIGURED (see Fig.14). The acquisition configuration can be
set as many time as required, provided that the system is not already acquiring (in such a case, stop
the acquisition to update the configuration).
To configure the level meter processing, use the Processing Settings panel:
- Obstacle Cancelation: Enable or disable obstacles effect cancelation. When the obstacles
cancelation is enabled, the effect of obstacles (for example, the contribution of the edge of
the ladle) is removed by the level meter processing. TIP: uncheck this flag when performing
level metering tests in an office environment, if you want to measure the distance of a not
moving target (for example a wall).
- Acquisition Average: Set an average factor for the acquisition repetition frequency (i.e. the
radar pulse repetition frequency). Measure update frequency (i.e. the frequency for the level
measure refresh) will be given by the radar pulse repetition frequency (set by selecting an
Acquisition speed value in the Acquisition Configuration panel), divided by this value.