ThesoftwarecanbefoundinmenuTools->”Pt100CalibrationTool”of
theloggerclientprogram.Belowyoundascreenshotofthetool.
Inthisformyousetthefollowingelds:
Current gain:
The current gain parameter setting used to measure the reference
temperatures.
Current offset:
The current offset parameter setting used to measure the reference
temperatures.
Reference Temperature 1:
The lower reference temperature value in degrees Celsius
Measured Temperature 1:
The measured temperature value in degrees Celsius
Reference Temperature 2:
The upper reference temperature value in degrees Celsius
Measured Temperature 2:
The measured temperature value in degrees Celsius
Whentheparametersareset,youndinthetwolowereldsthepara-
meters to set in the PRO-PT100 device:
Calculated gain as parameter 1.
The calculated offset as parameter 2.
After these parameters are set, you can verify if the sensor is indicating
the correct temperature for each reference. If needed you can repeat
the same procedure again.