Operation Manual

ARGUS
A
RGUS 125 103
13.1 Start the automatic test
In this case, the ARGUS executes a test sequence automatically. The required parameters
(e.g. measurement time and error level for the BERT see page 126) should be checked
before the automatic test series is begun.
Using the intec software, WINplus or WINanalyse, the test results can also be saved on a
PC, where they can be presented in a comprehensive measurement report and printed.
The ARGUS automatically performs the following sequence of single tests:
On a BRI or U-interface access
(ARGUS in TE Mode)
On a PRI access
(ARGUS in TE Mode)
- Status - Status
- Level measuring - Service tests
- Service tests - BERT in an extended call to oneself
- BERT in an extended call to oneself - Supplementary Services Test
(Suppl.serv.test)
- Supplementary Services Test
(Suppl.serv.test)
- CF Interrogation (Call Diversions)
- CF Interrogation (Call Diversions) - X.31 test
- MSN Interrogation - Read the L1 counter and
measurement time
- X.31 test
On a BRI or U-interface leased line On a PRI leased line (permanent circuit)
- Status - Status
- Level measuring - BERT in end-to-end mode (e.g. with a
loopbox on the remote end)
- Service tests - Read the L1 counter and
measurement time
- BERT in end-to-end mode (e.g. with a
loopbox on the remote end)