Technical data

15 Configuration
120
ARGUS 3u plus/3u NT
Settings for the BERT:
Display
Name on the
ARGUS
Comment
BERT -
duration
You can use the keypad to enter
measurement times ranging from 1
minute (default setting) to 99 hours
and 59 minutes (= 99:59).
If the time is set to 00:00
(=BERT with unlimited
measurement time), the BERT
will not stop automatically. In
this case, the user must
terminate the BERT by pressing
the < ABORT> softkey.
Error level This is the level used to evaluate
whether the BERT had an
"acceptable" bit error rate.
If the BERT has a bit error rate, which
exceeds this error level, the ARGUS
will display a "NO" as the test result.
Using the keypad, this parameter can
be set to any value from 01 (= 10
-01
) to
99 ( = 10
-99
).
The default threshold (error level) is
10
-06
(1E-06). That means that, in the
event that the bit error rate is less than
10
-06
(one error in 10
6
= 1,000,000
sent bits), the bit error test will be
evaluated as OK.
HRX value The HRX setting
(Hypothetical Reference Connections,
see the ITU-T G.821):
Using the keypad, you can enter a
value ranging from 0 to 100 %.